Effectiveness analysis of filters used with radon detectors under extreme environmental conditions for long-term exposures
View/Open
Cita com:
hdl:2117/178273
Document typeArticle
Defense date2015-02-04
PublisherElsevier
Rights accessOpen Access
Except where otherwise noted, content on this work
is licensed under a Creative Commons license
:
Attribution-NonCommercial-NoDerivs 3.0 Spain
Abstract
Active and passive radon detectors have been exposed with different filter configurations at the INTE radon chamber controlled conditions. Correction factors and delay times of the radon diffusion through each filter have been determined. Additionally, some of the studied filter/detector configurations have been used to measure radon in several workplaces and outdoor sites under real extreme environmental conditions. Analysis of these detectors showed partial degradation, so used filters seem not to be protective enough for long-term exposures.
CitationMoreno, V. [et al.]. Effectiveness analysis of filters used with radon detectors under extreme environmental conditions for long-term exposures. "Physics Procedia", 4 Febrer 2015, vol. 80, p. 113-116.
ISSN1875-3892
Files | Description | Size | Format | View |
---|---|---|---|---|
Moreno_et_al_2015.pdf | 240,0Kb | View/Open |