Comparison of two cardiac output monitors, qCO and LiDCO, during general anesthesia

View/Open
Document typeArticle
Defense date2016-08-09
Rights accessOpen Access
Except where otherwise noted, content on this work
is licensed under a Creative Commons license
:
Attribution-NonCommercial-NoDerivs 3.0 Spain
Abstract
Background: Optimization of cardiac output (CO) has been evidenced to reduce postoperative complications and to expedite the recovery. Likewise, CO and other dynamic cardiac parameters can describe the systemic blood flow and tissue oxygenation state and can be useful in different clinical fields. This study aimed to validate the qCO monitor (Quantium Medical, Barcelona, Spain), a new device to estimate CO and other related parameters in a continuous, fully non-invasive way using advanced digital signal processing of impedance cardiography.
Methods: The LiDCOrapidv2 (LiDCO Ltd, London, UK) was used to compare the performance of the qCO in 15 patients during major surgery under general anesthesia. Full surgeries were recorded and cardiac output obtained by both devices was compared by using correlation and Bland-Altman analysis.
Results: The Bland-Altman analysis showed sufficient agreement with a mean bias of -0.03 ± 0.71 L/min.
Conclusions: The findings showed that both systems offered comparable values and thus the non-invasive measurement of CO with qCO is a promising, feasible method. Further investigation will be required to validate this new device against calibrated devices and outcome studies would also be highly recommended.
CitationEscrivá, J., Gabarrón, E., Fort, P., Pan, Y., Shengjin , G., Liao, Q., Vallverdu, M. Comparison of two cardiac output monitors, qCO and LiDCO, during general anesthesia. "Mathews journal of anesthesia", 9 Agost 2016, vol. 1, núm. 1, p. 1-8.
Publisher versionhttp://www.mathewsopenaccess.com/PDF/Anesthesia/M_J_Anes_1_1_002.pdf