Applications of longitudinal and transversal bioimpedance measurements in peritoneal dialysis at 50 kHz
View/Open
1742-6596_224_1_012138-LEXA.pdf (499,2Kb) (Restricted access)
Request copy
Què és aquest botó?
Aquest botó permet demanar una còpia d'un document restringit a l'autor. Es mostra quan:
- Disposem del correu electrònic de l'autor
- El document té una mida inferior a 20 Mb
- Es tracta d'un document d'accés restringit per decisió de l'autor o d'un document d'accés restringit per política de l'editorial
Cita com:
hdl:2117/9232
Document typeConference report
Defense date2010
Rights accessRestricted access - publisher's policy
Except where otherwise noted, content on this work
is licensed under a Creative Commons license
:
Attribution-NonCommercial-NoDerivs 3.0 Spain
Abstract
More relevant information of the fluid changes in peritoneal dialysis (PD) might be
obtained with segmental bioimpedance measurements rather than whole-body measurement,
who hidden information of body composition. Whole-body and segmental bioimpedance
measurements were obtained using 5 configurations (whole-body or right-side (RS),
longitudinal-leg (L-LEG), longitudinal-abdomen (L-AB), transversal-abdomen (T-AB), and
transversal-leg (T-LEG)) in 20 patients: 15 males (56.5 ± 9.4 yr, 24.2 ± 4.2 kg/m2) and 5
females (58.4 ± 7.1 yr, 28.2 ± 5.9 kg/m2) in peritoneal dialysis (PD). The aim of this study is to
analyze the relationship between whole-body, longitudinal-segmental (L-LEG and L-AB) and
transversal-segmental (TAB and TLEG) bioimpedance measurement at 50 kHz, with clinical
parameters of cardiovascular risk, dyslipidemia, nutrition and hydration. The Kolmogorov-
Smirnov test was used for the normality test of all variables. Longitudinal bioimpedance
parameters were normalized by the height of the patients. The Spearman correlation was used
to analyze the correlation between bioimpedance and clinical parameters. The statistical
significance was considered with P < 0.05. Transversal bioimpedance measurements have
higher correlation with clinical parameters than longitudinal measurements.
CitationNescolarde Selva, Lexa. [et al.]. Applications of longitudinal and transversal bioimpedance measurements in peritoneal dialysis at 50 kHz. A: XIVth International Conference on Electrical Bioimpedance. "XIVth International Conference on Electrical Bioimpedance". Gainesville-Florida, April 04-08: 2010, p. 1-4.