dc.date.issued2016-05
dc.identifier.citationAmer, M., Chavez-Dominguez, J. A., Garcia, M., Salazar, J., Turo, A. Quartz crystal microbalance holder design for on-line sensing in liquid applications. "International Journal of Electrical, Computer, Energetic, Electronic and Communication Engineering", Maig 2016, vol. 10, núm. 5, p. 654-657.
dc.identifier.issn2010-376X
dc.identifier.urihttp://hdl.handle.net/2117/90555
dc.description.abstractIn this paper, the design of a QCM sensor for liquid media measurements in vertical position is described. A rugged and low-cost proof holder has been designed, the cost of which is significantly lower than those of traditional commercial holders. The crystal is not replaceable but it can be easily cleaned. Its small volume permits to be used by dipping it in the liquid with the desired location and orientation. The developed design has been experimentally validated by measuring changes in the resonance frequency and resistance of the QCM sensor immersed vertically in different calibrated aqueous glycerol solutions. The obtained results show a great agreement with the Kanazawa theoretical expression. Consequently, the designed QCM sensor would be appropriate for sensing applications in liquids, and might take part of a future on-line multichannel low-cost QCM-based measurement system.
dc.format.extent4 p.
dc.language.isoeng
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/3.0/es/
dc.subjectÀrees temàtiques de la UPC::Enginyeria electrònica::Instrumentació i mesura::Sensors i actuadors
dc.subject.lcshResonators
dc.subject.otherHolder design
dc.subject.otherliquid-media measurements
dc.subject.othermulti-channel measurements
dc.subject.otherQCM
dc.titleQuartz crystal microbalance holder design for on-line sensing in liquid applications
dc.typeArticle
dc.subject.lemacRessonadors
dc.contributor.groupUniversitat Politècnica de Catalunya. GSS - Grup Sistemes Sensors
dc.description.peerreviewedPeer Reviewed
dc.relation.publisherversionhttp://waset.org/publications/10004502/quartz-crystal-microbalance-holder-design-for-on-line-sensing-in-liquid-applications
dc.rights.accessOpen Access
local.identifier.drac18896618
dc.description.versionPostprint (published version)
dc.relation.projectidinfo:eu-repo/grantAgreement/MINECO//DPI2013-46915-C2-2-R/ES/DETECCION, CONTROL Y ACTUACION SOBRE BIOFILMS CON TRANSDUCTORES Y MATERIALES PIEZOELECTRICOS/
local.citation.authorAmer, M.; Chavez-Dominguez, J. A.; Garcia, M.; Salazar, J.; Turo, A.
local.citation.publicationNameInternational Journal of Electrical, Computer, Energetic, Electronic and Communication Engineering
local.citation.volume10
local.citation.number5
local.citation.startingPage654
local.citation.endingPage657