Low dose response results and detection limits of the EURADOS intercomparison IC2021area for passive H*(10) area dosemeters

Thumbnail
View/Open
Article en format word (649,8Kb) (Restricted access)   Request copy 

Què és aquest botó?

Aquest botó permet demanar una còpia d'un document restringit a l'autor. Es mostra quan:

  • Disposem del correu electrònic de l'autor
  • El document té una mida inferior a 20 Mb
  • Es tracta d'un document d'accés restringit per decisió de l'autor o d'un document d'accés restringit per política de l'editorial
 
10.1016/j.radmeas.2023.106935
 
  View Usage Statistics
  LA Referencia / Recolecta stats
Cita com:
hdl:2117/388373

Show full item record
Hranitzky, Christian
Naber, Christian
Aslan, Julia
Duch Guillen, María AmorMés informacióMés informacióMés informació
Haninger, Thomas
Kneževic, Zeljka
Document typeArticle
Defense date2023-04-01
Rights accessRestricted access - publisher's policy (embargoed until 2024-12)
Attribution-NonCommercial-NoDerivs 4.0 International
Except where otherwise noted, content on this work is licensed under a Creative Commons license : Attribution-NonCommercial-NoDerivs 4.0 International
Abstract
The EURADOS intercomparison IC2021area with its main focus on low dose measurements was carried out between May 2021 and April 2022 for passive H*(10) area dosimetry systems. Three measurement conditions were provided at locations of the Karlsruhe Institute of Technology: 3-months indoor, 3-months outdoor and 6-months outdoor. The challenge for the 65 participating dosimetry systems of this intercomparison was measuring additionally irradiated low dose 137Cs gamma radiation of 150 µSv and 300 µSv. Despite the comparably high background dose values between about 200 µSv and 1600 µSv, more than 90% of the resulting response values of the irradiated dosemeters were within the recommended ISO 14146 response limits. The calculated detection limits of the participating dosimetry systems were in most cases clearly below the desired value of 150 µSv of this intercomparison.
CitationHranitzky, C. [et al.]. Low dose response results and detection limits of the EURADOS intercomparison IC2021area for passive H*(10) area dosemeters. "Radiation measurements", 1 Abril 2023, vol. 163, núm. 106935. 
URIhttp://hdl.handle.net/2117/388373
DOI10.1016/j.radmeas.2023.106935
ISSN1350-4487
Publisher versionhttps://www.sciencedirect.com/science/article/abs/pii/S1350448723000367
Collections
  • INTE - Institut de Tècniques Energètiques - Articles de revista [280]
  View Usage Statistics

Show full item record

FilesDescriptionSizeFormatView
Hranitzky_2023.docxBlockedArticle en format word649,8Kb Microsoft Word 2007 Restricted access