Implant-abutment connections: influence of the design on the microgap and their fatigue and fracture behavior of dental implants
Visualitza/Obre
Article (1,286Mb) (Accés restringit)
Sol·licita una còpia a l'autor
Què és aquest botó?
Aquest botó permet demanar una còpia d'un document restringit a l'autor. Es mostra quan:
- Disposem del correu electrònic de l'autor
- El document té una mida inferior a 20 Mb
- Es tracta d'un document d'accés restringit per decisió de l'autor o d'un document d'accés restringit per política de l'editorial
Cita com:
hdl:2117/27711
Tipus de documentArticle
Data publicació2014-04-10
Condicions d'accésAccés restringit per política de l'editorial
Llevat que s'hi indiqui el contrari, els
continguts d'aquesta obra estan subjectes a la llicència de Creative Commons
:
Reconeixement-NoComercial-SenseObraDerivada 3.0 Espanya
Abstract
Microgap between implant and abutment can produce biological and mechanical problems such as peri-implantitis and/or fatigue failures. The aim of this study
was to evaluate microgap size and fatigue behavior of external and internal connections. In both systems the torque to tighten the abutment screw of single crown abutments was 45 Ncm. Fifty implants for each connection type were studied. One subgroup (n=5) was used by the observation and evaluation of the microgap, other (n=5) was tested for fracture strength and the other (n=40) was
subjected to dynamic loading. The internal connection presents a lower microgap than the external ones. From
fatigue results, the external hexagon interface showed superior result compared to the internal hexagon interfaces.
The tolerances in the internal connections are better and
favour the fatigue behavior but this factor alone is not sufficient to improve the fatigue response in relation to the
external connections when the screw is subjected at the same torque. The external system presents a higher value
of the area than the internal and it produces a better load distribution. Microgaps and mechanical properties are very
important for the long-term behavior of the dental implants and these aspects should be known by the implantologists
CitacióGil, F.J. [et al.]. Implant-abutment connections: influence of the design on the microgap and their fatigue and fracture behavior of dental implants. "Journal of materials science. Materials in medicine", 10 Abril 2014, vol. 25, núm. 7, p. 1825-1830.