One-point microwave radiometer calibration

View/Open
Cita com:
hdl:2117/191571
Document typeArticle
Defense date2020-03-01
Rights accessOpen Access
Abstract
A method for internally calibrating microwave total power radiometers by using only one level of noise injection is presented. It is based on having a previous accurate characterization of the receiver noise temperature, which used de facto as a second calibration standard. The method proves to be at least equivalent to the classical two level, as demonstrated through their intercomparison using the data provided by the Microwave Imaging Radiometer using Aperture Synthesis (MIRAS) on board the European Space Agency Soil Moisture and Ocean Salinity (SMOS) Satellite. The long-term stability in terms of retrieved brightness temperature using both methods has similar trends with a small advantage for the one-point approach proposed here.
Description
© 2020 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes,creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works.
CitationCorbella, I. [et al.]. One-point microwave radiometer calibration. "IEEE Geoscience and Remote Sensing Letters", 1 Març 2020, vol. 17, núm. 3, p. 461-463.
ISSN1545-598X
Publisher versionhttps://ieeexplore.ieee.org/document/8752053
Files | Description | Size | Format | View |
---|---|---|---|---|
08752053.pdf | 504,1Kb | View/Open |
All rights reserved. This work is protected by the corresponding intellectual and industrial
property rights. Without prejudice to any existing legal exemptions, reproduction, distribution, public
communication or transformation of this work are prohibited without permission of the copyright holder