A method for accurate zero calibration of asymmetric jaws in single-isocenter half-beam techniques

Thumbnail
View/Open
A method.pdf (349,3Kb) (Restricted access)   Request copy 

Què és aquest botó?

Aquest botó permet demanar una còpia d'un document restringit a l'autor. Es mostra quan:

  • Disposem del correu electrònic de l'autor
  • El document té una mida inferior a 20 Mb
  • Es tracta d'un document d'accés restringit per decisió de l'autor o d'un document d'accés restringit per política de l'editorial
Share:
 
 
10.1118/1.4773314
 
  View Usage Statistics
Cita com:
hdl:2117/18727

Show full item record
Hernandez Espinosa, Victor
Sempau Roma, JosepMés informacióMés informacióMés informació
Abella, R.
Romero Lopez, Matilde
Lopez Perez, Marcedes
Artigues, Manel
Arenas, M.
Document typeArticle
Defense date2013-02
Rights accessRestricted access - publisher's policy
Attribution-NonCommercial-NoDerivs 3.0 Spain
Except where otherwise noted, content on this work is licensed under a Creative Commons license : Attribution-NonCommercial-NoDerivs 3.0 Spain
Abstract
Purpose: To present a practical method for calibrating the zero position of asymmetric jaws that provides higher accuracy at the central axis and improves dose homogeneity in the abutting region of half-beams. Methods: Junction doses were measured for each asymmetric jaw using the double-exposure technique and electronic portal imaging devices. The junction dose was determined as a function of jaw position. The shift in the zero jaw position (or in its corresponding potentiometer readout) required to correct for the measured junction dose could thus be obtained. The jaw calibration was then modified to introduce the calculated shift and therefore achieve an accurate zero position in order to provide a relative junction dose that was as close to zero as possible. Results: All the asymmetric jaws from four medical linear accelerators were calibrated with the new calibration procedure. Measured relative junction doses at gantry 0° were reduced from a maximum of ±40% to a maximum of ±8% for all the jaws in the four considered accelerators. These results were valid for 6 MV and 18 MV photon beams and for any combination of asymmetric jaws set to zero. The calibration was stable over a long period of time; therefore, the need for recalibrating is seldom necessary. Conclusions: Accurate calibration of the zero position of the jaws is feasible in current medical linear accelerators. The proposed procedure is fast and it improves dose homogeneity at the junction of half-beams, thus, allowing a more accurate and safer use of these techniques. © 2013 American Association of Physicists in Medicine.
CitationHernandez, V. [et al.]. A method for accurate zero calibration of asymmetric jaws in single-isocenter half-beam techniques. "Medical physics", Febrer 2013, vol. 40, núm. 2, p. 1-6. 
URIhttp://hdl.handle.net/2117/18727
DOI10.1118/1.4773314
ISSN0094-2405
Publisher versionhttp://www.scopus.com/record/display.url?eid=2-s2.0-84873586272&origin=inward&txGid=D24C552D0A97633B3A6ECBFF4FC06A45.WlW7NKKC52nnQNxjqAQrlA%3a7
Collections
  • GREENER - Grup de recerca d'estudis energètics i de les radiacions - Articles de revista [178]
  • INTE - Institut de Tècniques Energètiques - Articles de revista [275]
Share:
 
  View Usage Statistics

Show full item record

FilesDescriptionSizeFormatView
A method.pdfBlocked349,3KbPDFRestricted access

Browse

This CollectionBy Issue DateAuthorsOther contributionsTitlesSubjectsThis repositoryCommunities & CollectionsBy Issue DateAuthorsOther contributionsTitlesSubjects

© UPC Obrir en finestra nova . Servei de Biblioteques, Publicacions i Arxius

info.biblioteques@upc.edu

  • About This Repository
  • Contact Us
  • Send Feedback
  • Inici de la pàgina