dc.contributor.authorKollaard, Robert
dc.contributor.authorCarinou, Eleftheria
dc.contributor.authorGinjaume Egido, Mercè
dc.contributor.authorStruelens, L.
dc.contributor.otherUniversitat Politècnica de Catalunya. Institut de Tècniques Energètiques
dc.date.accessioned2020-04-03T08:52:41Z
dc.date.issued2019-12-31
dc.identifier.citationKollaard, R. [et al.]. How to establish an adequate system for eye lens dose monitoring: a proposal for typical workplaces. "Radiation protection dosimetry", 31 Desembre 2019, vol. 185, núm. 3, p. 296-302.
dc.identifier.issn0144-8420
dc.identifier.urihttp://hdl.handle.net/2117/183016
dc.description.abstractFor years, the dose limit of 150 mSv for occupational exposure of the lens of the eye to ionising radiation was rarely exceeded, and the dose to the eye was only monitored occasionally. With the national implementation of the European Basic Safety Standards in 2018, this dose limit was reduced to 20 mSv and the Member States are expected to implement an adequate system for the monitoring of category A workers. Where the system for monitoring the whole body dose is settled in most countries, this is not the situation for the lens of the eye. This article presents a system for eye lens dose monitoring, based on the particle type, energy, angle of incidence and geometry of the radiation field and the use of protective measures. The system provides recommendations for the adequate operational quantity and dosemeter position for some of the most relevant workplaces.
dc.format.extent7 p.
dc.language.isoeng
dc.rightsAttribution-NonCommercial-NoDerivs 3.0 Spain
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/3.0/es/
dc.subjectÀrees temàtiques de la UPC::Física
dc.subjectÀrees temàtiques de la UPC::Ciències de la visió
dc.subjectÀrees temàtiques de la UPC::Ciències de la visió::Instruments òptics i optomètrics
dc.subject.lcshLenses
dc.subject.lcshOphthalmic lenses
dc.subject.lcshRadiation dosimetry
dc.titleHow to establish an adequate system for eye lens dose monitoring: a proposal for typical workplaces
dc.typeArticle
dc.subject.lemacLents
dc.subject.lemacLents oftàlmiques
dc.subject.lemacRadiació--Dosimetria
dc.contributor.groupUniversitat Politècnica de Catalunya. DRM - Dosimetria i Radiofísica Mèdica
dc.identifier.doi10.1093/rpd/ncz009
dc.relation.publisherversionhttps://academic.oup.com/rpd/article/185/3/296/5315815
dc.rights.accessRestricted access - publisher's policy
local.identifier.drac27186967
dc.description.versionPostprint (author's final draft)
local.citation.authorKollaard, R.; Carinou, E.; Ginjaume, M.; Struelens, L.
local.citation.publicationNameRadiation protection dosimetry
local.citation.volume185
local.citation.number3
local.citation.startingPage296
local.citation.endingPage302