Mostra el registre d'ítem simple

dc.contributor.authorGarcia, J.
dc.contributor.authorDoce, Y.
dc.contributor.authorManuel, F.
dc.contributor.authorCastro, F.
dc.contributor.authorGoicolea, J.
dc.contributor.authorFernández, A.
dc.date.accessioned2019-07-26T08:14:53Z
dc.date.available2019-07-26T08:14:53Z
dc.date.issued2016
dc.identifier.citationGarcia, J. [et al.]. Estudio comparativo de diferentes técnicas de implantación de stents en bifurcaciones coronarias. "Revista internacional de métodos numéricos para cálculo y diseño en ingeniería", 2016, vol. 32, núm. 3.
dc.identifier.issn1886-158X
dc.identifier.issn0213-1315
dc.identifier.urihttp://hdl.handle.net/2117/166899
dc.description.abstractEl objetivo de este trabajo es el análisis comparativo de los cambios hemodinámicos locales producidos en una bifurcación coronaria por el implante de stents mediante 3 técnicas diferentes: stent simple en la arteria principal, stent simple en la arteria principal con orificio de comunicación con la arteria secundaria para el paso de un balón de inflado («kissing») y «culotte» . Para ello se ha utilizado la geometría ideal de una bifurcación coronaria a 45 y 90°. En primer lugar se simularon los modelos en régimen estacionario mediante el código comercial ANSYS Fluent, y a continuación se validaron con las medidas experimentales tomadas en el laboratorio mediante velocimetría de imagen de partículas (particle image velocimetry [PIV]). Una vez obtenida una buena concordancia de resultados entre ambos estudios en régimen permanente, se abordó el flujo pulsátil. Los parámetros que se consideraron para comparar las diferentes técnicas en este régimen fueron los esfuerzos cortantes y la caída de presión promediados en el tiempo, así como el índice de esfuerzos cortantes oscilatorios (oscillatory shear index [OSI]). Las conclusiones finales fueron que el stent simple presenta los peores resultados, en términos de caída de presión, y la técnica «kissing» , los mejores. Esto sería aplicable a los stents liberadores de fármacos, donde la reestenosis no es tan crucial como el restablecimiento del flujo sanguíneo. En cuanto al criterio basado en el valor menor de los esfuerzos cortantes (< 0,4 Pa) y un OSI elevado (> 0,1), la técnica «culotte» presenta el peor comportamiento. Esto sería aplicable a los stents sin recubrimiento, en que la mejor opción sería el stent simple.
dc.description.abstractThe objective of this work is to analyze the local hemodynamic changes caused in a coronary bifurcation by three different stenting techniques: simple stenting of the main vessel, simple stenting of the main vessel with kissing balloon in the side branch and culotte. To carry out this study an idealized geometry of a coronary bifurcation is used, and two bifurcation angles, 45° and 90°, are chosen as representative of the wide variety of real configurations. Both numerical simulations and experimental measurements are performed. First, steady simulations are carried out with the commercial code ANSYS Fluent, then, experimental measurements with PIV (Particle Image Velocimetry), obtained in the laboratory, are used to validate the numerical simulations. The steady computational simulations show a good overall agreement with the experimental data. Second, pulsatile flow is considered to take into account the transient effects. The time averaged wall shear stress, oscillatory shear index (OSI) and pressure drop obtained numerically are used to compare the behavior of the stenting techniques. As conclusions obtained, simple stenting technique shows the worst results in terms of pressure drop, and kissing the best ones. This criterion could be applied for DES (Drug Eluting Stents) where the restenosis is not as critical as the reestablishment of the blood flow. But considering the regions of low wall shear stress (< 0.4 Pa) and elevated OSI (> 0.1), the culotte procedure shows the worst performance and the simple stent, the best. This could be applied to BMS (Bare Metal Stents).
dc.language.isospa
dc.publisherUniversitat Politècnica de Catalunya. CIMNE
dc.rightsAttribution-NonCommercial-ShareAlike 3.0 Unported
dc.rightsAttribution-NonCommercial-NoDerivs 3.0 Spain
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/3.0/es/
dc.subjectÀrees temàtiques de la UPC::Matemàtiques i estadística::Anàlisi numèrica
dc.subject.lcshNumerical analysis
dc.subject.otherStent
dc.subject.otherBifurcación coronaria
dc.subject.otherDinámica de fluidos computacional
dc.subject.otherEsfuerzo cortante
dc.subject.otherCoronary bifurcation
dc.subject.otherComputational fluid dynamics
dc.subject.otherWall shear stress
dc.titleEstudio comparativo de diferentes técnicas de implantación de stents en bifurcaciones coronarias
dc.title.alternativeEvaluation of different stenting techniques in coronary bifurcations
dc.typeArticle
dc.subject.lemacAnàlisi numèrica
dc.identifier.doi10.1016/j.rimni.2015.04.003
dc.description.peerreviewedPeer Reviewed
dc.rights.accessOpen Access
local.citation.publicationNameRevista internacional de métodos numéricos para cálculo y diseño en ingeniería
local.citation.volume32
local.citation.number3


Fitxers d'aquest items

Thumbnail

Aquest ítem apareix a les col·leccions següents

Mostra el registre d'ítem simple