dc.contributor.author | González-Mora, V. A. |
dc.contributor.author | Hoffmann, Michael |
dc.contributor.author | Stroosnijder, R. |
dc.contributor.author | Gil Mur, Francisco Javier |
dc.contributor.other | Universitat Politècnica de Catalunya. Departament de Ciència dels Materials i Enginyeria Metal·lúrgica |
dc.date.accessioned | 2012-02-15T12:49:47Z |
dc.date.available | 2012-02-15T12:49:47Z |
dc.date.created | 2011 |
dc.date.issued | 2011 |
dc.identifier.citation | González-Mora, V.A. [et al.]. The role of hardness and roughness on the wear of different CoCrMo counterfaces on UHMWPE for artificial joints. "Journal of Biomedical Science and Engineering", 2011, vol. 4, núm. 10, p. 651-656. |
dc.identifier.issn | 1423-0127 |
dc.identifier.uri | http://hdl.handle.net/2117/15160 |
dc.description.abstract | Wear tests were carried out to study the effect of the hardness and roughness with various counterface materials on UHMWPE wear behaviour. The materials used as counterfaces were based on varieties of CoCrMo: 1) forged (hand-polished) CoCrMo, 2) forged (mass-finished) CoCrMo, and 3) cast (mass-finished) CoCrMo. Additionally, two coatings were proposed: 1) a CoCrMo coating applied to the forged CoCrMo alloy by means of physical vapour deposition (PVD), and 2) a ZrO2 coating applied to the forged CoCrMo alloy by means of plasma-assisted chemical vapour deposition (PACVD). The reciprocating pin-on-flat (RPOF) device for pin-on-disk wear testing was used for this study. The worn surfaces were observed using optical, atomic force and scanning electron microscopes. |
dc.format.extent | 6 p. |
dc.language.iso | eng |
dc.publisher | Springer |
dc.rights | Attribution-NonCommercial-NoDerivs 3.0 Spain |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/3.0/es/ |
dc.subject | Àrees temàtiques de la UPC::Enginyeria dels materials |
dc.subject.lcsh | Artificial joints |
dc.title | The role of hardness and roughness on the wear of different CoCrMo counterfaces on UHMWPE for artificial joints |
dc.type | Article |
dc.subject.lemac | Articulacions artificials |
dc.contributor.group | Universitat Politècnica de Catalunya. BBT - Biomaterials, Biomecànica i Enginyeria de Teixits |
dc.identifier.doi | 10.4236/jbise.2011.410081 |
dc.description.peerreviewed | Peer Reviewed |
dc.relation.publisherversion | http://www.scirp.org/journal/PaperInformation.aspx?PaperID=7908&JournalID=30 |
dc.rights.access | Open Access |
local.identifier.drac | 9597283 |
dc.description.version | Postprint (published version) |
local.citation.author | González-Mora, V.A.; Hoffman, M.; Stroosnijder, R.; Gil, F.J. |
local.citation.publicationName | Journal of Biomedical Science and Engineering |
local.citation.volume | 4 |
local.citation.number | 10 |
local.citation.startingPage | 651 |
local.citation.endingPage | 656 |