Mostra el registre d'ítem simple

dc.contributor.authorRosell Ferrer, Francisco Javier
dc.contributor.authorWebster, John G.
dc.contributor.otherUniversitat Politècnica de Catalunya. Departament d'Enginyeria Electrònica
dc.date.accessioned2007-11-20T14:53:23Z
dc.date.available2007-11-20T14:53:23Z
dc.date.created1995-03-01
dc.date.issued1995-03-01
dc.identifier.citationRosell, J; Webster, GJ. Signal-to-motion artifact ratio versus frequency for impedance pneumograhy. IEEE Transactions on Biomedical Engineering, 1995, vol. 42, núm.3, p. 321-323
dc.identifier.issn0018-9294
dc.identifier.urihttp://hdl.handle.net/2117/1348
dc.description.abstractWe measured transthoracic impedance between 12.5 and 185 kHz in nine adults. We used a system with two impedance channels, both simultaneously detecting the real part of impedance at two different frequencies. We used only two electrodes in the midaxillary line, connecting both channels in parallel. The amplitude relation between the two channels was measured for different maneuvers and frequencies. Results show for normal breathing an increase of the signal of 20% and a decrease in motion artifacts from 12.5 to 185 kHz. We conclude that, for the maneuvers studied, it is better to work at higher frequencies than the ones commonly used. Also, we suggest a method to further increase the signal-to-motion artifact ratio based on measurement at two frequencies.
dc.format.extent321-323
dc.language.isoeng
dc.publisherIEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
dc.subjectÀrees temàtiques de la UPC::Enginyeria biomèdica::Electrònica biomèdica
dc.subject.lcshElectrodiagnosis
dc.subject.lcshElectrophysiology
dc.subject.otherBioelectric phenomena
dc.subject.otherBiomedical measurement
dc.subject.otherElectric impedance measurement
dc.subject.otherMedical signal processing
dc.subject.otherSignal-to-motion artifact radio
dc.subject.otherImpedance pneumography
dc.subject.otherMotion artifacts
dc.subject.otherManeuvers
dc.subject.otherMidaxilary line
dc.subject.otherNormal breathing
dc.titleSignal-to-motion artifact ratio versus frequency for impedance pneumography
dc.typeArticle
dc.subject.lemacElectrodiagnòstic
dc.subject.lemacElectrofisiologia -- Processament de dades
dc.contributor.groupUniversitat Politècnica de Catalunya. IEB - Instrumentació Electrònica i Biomèdica
dc.description.peerreviewedPeer Reviewed
dc.rights.accessOpen Access
dc.relation.projectidcttDGICYT PB92/0892
dc.relation.projectidcttDGICYT 93/197
local.personalitzacitaciotrue


Fitxers d'aquest items

Thumbnail

Aquest ítem apareix a les col·leccions següents

Mostra el registre d'ítem simple