dc.contributor.author | Gultresa Colomer, Joan |
dc.contributor.author | Llansana, Jordi |
dc.contributor.author | Roig Costa, Maria |
dc.contributor.author | Ginjaume Egido, Mercè |
dc.contributor.other | Universitat Politècnica de Catalunya. Institut de Tècniques Energètiques |
dc.date.accessioned | 2019-02-11T10:37:45Z |
dc.date.available | 2019-02-11T10:37:45Z |
dc.date.issued | 2016-09-01 |
dc.identifier.citation | Gultresa, J. [et al.]. A practical approach to perform the isotropy test for extremity dosemeters. "Radiation protection dosimetry", 1 Setembre 2016, vol. 170, núm. 1-4, p. 95-99. |
dc.identifier.issn | 0144-8420 |
dc.identifier.uri | http://hdl.handle.net/2117/128867 |
dc.description.abstract | The requirements for determining extremity dosemeter performance have evolved over the past decade. In 2010, the Spanish Nuclear Safety Council (CSN) adopted a national protocol with performance requirements based on ISO 12794. Because of the lack of symmetry of ISO 4037-3 calibration phantoms, the isotropy test set up in the sagittal plane presented several challenges both for individual monitoring services and for calibration labs. This article proposes and validates a practical approach to reduce the number of irradiations. Results show that wrist and ring dosemeters in this study meet the ISO 12794 requirements for the isotropy test. However, additional studies would be needed to verify the newer IEC 62387 Standard. |
dc.format.extent | 5 p. |
dc.language.iso | eng |
dc.rights | Attribution-NonCommercial-NoDerivs 3.0 Spain |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/3.0/es/ |
dc.subject | Àrees temàtiques de la UPC::Física |
dc.subject | Àrees temàtiques de la UPC::Física::Electromagnetisme |
dc.subject | Àrees temàtiques de la UPC::Ciències de la salut::Medicina::Diagnòstic per la imatge |
dc.subject.lcsh | Radiation dosimetry |
dc.subject.lcsh | Medical physics |
dc.title | A practical approach to perform the isotropy test for extremity dosemeters |
dc.type | Article |
dc.subject.lemac | Radiació--Dosimetria |
dc.subject.lemac | Física mèdica |
dc.contributor.group | Universitat Politècnica de Catalunya. DRM - Dosimetria i Radiofísica Mèdica |
dc.identifier.doi | 10.1093/rpd/ncv530 |
dc.relation.publisherversion | https://academic.oup.com/rpd/article-abstract/170/1-4/95/2223851/A-PRACTICAL-APPROACH-TO-PERFORM-THE-ISOTROPY-TEST?redirectedFrom=fulltext |
dc.rights.access | Open Access |
local.identifier.drac | 19810679 |
dc.description.version | Postprint (author's final draft) |
local.citation.author | Gultresa, J.; Llansana, J.; Roig, M.; Ginjaume, M. |
local.citation.publicationName | Radiation protection dosimetry |
local.citation.volume | 170 |
local.citation.number | 1-4 |
local.citation.startingPage | 95 |
local.citation.endingPage | 99 |