dc.identifier.citationGultresa, J. [et al.]. A practical approach to perform the isotropy test for extremity dosemeters. "Radiation protection dosimetry", 1 Setembre 2016, vol. 170, núm. 1-4, p. 95-99.
dc.identifier.issn0144-8420
dc.identifier.urihttp://hdl.handle.net/2117/128867
dc.description.abstractThe requirements for determining extremity dosemeter performance have evolved over the past decade. In 2010, the Spanish Nuclear Safety Council (CSN) adopted a national protocol with performance requirements based on ISO 12794. Because of the lack of symmetry of ISO 4037-3 calibration phantoms, the isotropy test set up in the sagittal plane presented several challenges both for individual monitoring services and for calibration labs. This article proposes and validates a practical approach to reduce the number of irradiations. Results show that wrist and ring dosemeters in this study meet the ISO 12794 requirements for the isotropy test. However, additional studies would be needed to verify the newer IEC 62387 Standard.
dc.format.extent5 p.
dc.language.isoeng
dc.rightsAttribution-NonCommercial-NoDerivs 3.0 Spain
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/3.0/es/
dc.subjectÀrees temàtiques de la UPC::Física
dc.subjectÀrees temàtiques de la UPC::Física::Electromagnetisme
dc.subjectÀrees temàtiques de la UPC::Ciències de la salut::Medicina::Diagnòstic per la imatge
dc.subject.lcshRadiation dosimetry
dc.subject.lcshMedical physics
dc.titleA practical approach to perform the isotropy test for extremity dosemeters
dc.typeArticle
dc.subject.lemacRadiació--Dosimetria
dc.subject.lemacFísica mèdica
dc.contributor.groupUniversitat Politècnica de Catalunya. DRM - Dosimetria i Radiofísica Mèdica
dc.identifier.doi10.1093/rpd/ncv530
dc.relation.publisherversionhttps://academic.oup.com/rpd/article-abstract/170/1-4/95/2223851/A-PRACTICAL-APPROACH-TO-PERFORM-THE-ISOTROPY-TEST?redirectedFrom=fulltext
dc.rights.accessOpen Access
local.identifier.drac19810679
dc.description.versionPostprint (author's final draft)
local.citation.authorGultresa, J.; Llansana, J.; Roig, M.; Ginjaume, M.
local.citation.publicationNameRadiation protection dosimetry
local.citation.volume170
local.citation.number1-4
local.citation.startingPage95
local.citation.endingPage99


Files in this item