Calibration of Raman lidar water vapor mixing ratio measurements using zenithal measurements of diffuse sunlight and a radiative transfer model
View/Open
Cita com:
hdl:2117/126308
Document typeArticle
Defense date2018-07-20
PublisherInstitute of Electrical and Electronics Engineers (IEEE)
Rights accessOpen Access
All rights reserved. This work is protected by the corresponding intellectual and industrial
property rights. Without prejudice to any existing legal exemptions, reproduction, distribution, public
communication or transformation of this work are prohibited without permission of the copyright holder
Abstract
Among the different techniques available for measuring the atmospheric water vapor content, Raman lidars stand out as accurate instruments providing detailed profiles with high temporal and altitude resolution. Their principle is based on obtaining the range-resolved ratio of the lidar signals corresponding to Raman returns from water vapor and nitrogen molecules, which is proportional to the water vapor mixing ratio. To do this, it is necessary to determine a calibration factor, specific of each lidar instrument. A method for obtaining this parameter, based on zenith measurements of diffuse sunlight, on Raman scattering models and on simulations, using a radiative transfer model, to estimate sky radiances at the wavelengths of interest, has been applied to the lidar system of Universitat Politècnica de Catalunya (UPC; Technical University of Catalonia, Barcelona, Spain). A set of calibrations, performed between 2016 and 2017, has permitted assessing the calibration procedure and analyzing the stability of the calibration factor in the UPC instrument. Results show that although the calibration factor can remain stable for long periods of time, it can suffer sudden variations that make indispensable to implement a convenient and reliable procedure to perform regular calibrations. We show that the method, which can be applied to any lidar with water vapor and nitrogen Raman channels, can completely dispense with radiosonde data. The calibration method is validated by comparison with simultaneous radiosonde water vapor measurements. Limitations of radiosondes for validating--and eventually calibrating--water vapor Raman lidars have been revealed.
Description
This is a postprint (author final draft) version of article that has been accepted for publication. A fully version can be found at: https://doi.org/10.1109/TGRS.2018.2851064
© 2018 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works.
CitationMuñoz, C., Comeron, A., Sicard, M., Barragan, R., Garcia, D., Rodriguez-Gomez, A., Rocadenbosch, F., Granados, M. Calibration of Raman lidar water vapor mixing ratio measurements using zenithal measurements of diffuse sunlight and a radiative transfer model. "IEEE transactions on geoscience and remote sensing", 20 Juliol 2018, vol. 56, núm. 12, p. 7405-7414.
ISSN0196-2892
Publisher versionhttps://ieeexplore.ieee.org/document/8415748
Files | Description | Size | Format | View |
---|---|---|---|---|
08415748.pdf | 3,139Mb | View/Open |