Mostra el registre d'ítem simple

dc.contributor.authorMillán Blasco, Oliver
dc.contributor.authorSalazar Soler, Jorge
dc.contributor.authorChávez Domínguez, Juan Antonio
dc.contributor.authorTuró Peroy, Antonio
dc.contributor.authorGarcía Hernández, Miguel J.
dc.contributor.otherUniversitat Politècnica de Catalunya. Departament d'Enginyeria Electrònica
dc.date.accessioned2017-11-24T09:27:18Z
dc.date.available2017-11-24T09:27:18Z
dc.date.issued2017-11
dc.identifier.citationMillan, O., Salazar, J., Chavez-Dominguez, J. A., Turo, A., Garcia, M. Zero-flow offset variation in ultrasonic clamp-on flowmeters due to inhomogeneity and nonlinearity of pipe materials. "IEEE transactions on instrumentation and measurement", Novembre 2017, vol. 66, núm. 11, p. 2845-2851.
dc.identifier.issn0018-9456
dc.identifier.urihttp://hdl.handle.net/2117/111146
dc.description.abstractThe objective of this paper is to measure the zero-flow offset variation in ultrasonic Clamp-on flowmeters produced by using pipes made of materials with a nonhomogeneous and nonlinear behavior. The evaluation of this phenomenon has been tested by the experimental measurement, the experiments consist of measuring the variation in the flow measurement when in ``no-flow condition,'' the transducers are moved along the pipe. Experimental results reveal that the reciprocity criterion is not fulfilled when nonhomogeneous and nonlinear pipe materials are used. Consequently, in these types of pipes materials, it is necessary to calibrate the offset when transducers are installed, and probably will need recalibration if the transducers are displaced to another position on the pipe. However, in installations, where it is not possible to stop flow in order to calibrate, the flowmeter will have an error in the measured value and this error will depend on the acoustic characteristics of the pipe material.
dc.format.extent7 p.
dc.language.isoeng
dc.subjectÀrees temàtiques de la UPC::Enginyeria electrònica
dc.subject.lcshTransducers
dc.subject.lcshCalibration
dc.subject.lcshAcoustics
dc.subject.lcshSteel
dc.subject.lcshLiquids
dc.subject.otherAcoustic measurements
dc.subject.otherAcoustics
dc.subject.otherCalibration
dc.subject.otherDesviation of zero-flow measurement
dc.subject.otherinhomogeneity and nonlinearity of material
dc.subject.otherLiquids
dc.subject.otherSteel
dc.subject.otherTransducers
dc.subject.otherultrasonic Clamp-on flowmeters.
dc.subject.otherUltrasonic variables measurement
dc.titleZero-flow offset variation in ultrasonic clamp-on flowmeters due to inhomogeneity and nonlinearity of pipe materials
dc.typeArticle
dc.subject.lemacTransductors
dc.subject.lemacCalibratge
dc.subject.lemacAcer
dc.subject.lemacLíquids
dc.contributor.groupUniversitat Politècnica de Catalunya. GSS - Grup Sistemes Sensors
dc.identifier.doi10.1109/TIM.2017.2731018
dc.description.peerreviewedPeer Reviewed
dc.relation.publisherversionhttp://ieeexplore.ieee.org/document/8007251/
dc.rights.accessRestricted access - publisher's policy
local.identifier.drac21554046
dc.description.versionPostprint (author's final draft)
local.citation.authorMillan, O.; Salazar, J.; Chavez-Dominguez, J. A.; Turo, A.; Garcia, M.
local.citation.publicationNameIEEE transactions on instrumentation and measurement
local.citation.volume66
local.citation.number11
local.citation.startingPage2845
local.citation.endingPage2851