Ag nanoparticles drop-casting modification of screen-printed electrodes for the simultaneous voltammetric determination of Cu(II) and Pb(II)

View/Open
Document typeArticle
Defense date2017-06-21
PublisherMDPI AG
Rights accessOpen Access
Abstract
A new silver nanoparticle modified screen-printed electrode was developed and applied to the simultaneous determination of Pb(II) and Cu(II). Two different types of silver nanoparticles with different shapes and sizes, Ag nanoseeds and Ag nanoprisms, were microscopically characterized and three different carbon substrates, graphite, graphene and carbon nanofibers, were tested. The best analytical performance was achieved for the combination of Ag nanoseeds with a carbon nanofiber modified screen-printed electrode. The resulting sensor allowed the simultaneous determination of Pb(II) and Cu(II) at trace levels and its applicability to natural samples was successfully tested with a groundwater certified reference material, presenting high reproducibility and trueness.
CitationPérez, C., Bastos-Arrieta, J., Serrano, N., Díaz, J., Ariño, C., Pablo, J., Esteban, M. Ag nanoparticles drop-casting modification of screen-printed electrodes for the simultaneous voltammetric determination of Cu(II) and Pb(II). "Sensors", 21 Juny 2017, vol. 17, núm. 6.
ISSN1424-8220
Publisher versionhttp://www.mdpi.com/1424-8220/17/6/1458
Files | Description | Size | Format | View |
---|---|---|---|---|
sensors-17-01458.pdf | 3,216Mb | View/Open |
Except where otherwise noted, content on this work
is licensed under a Creative Commons license
:
Attribution-NonCommercial-NoDerivs 3.0 Spain