Optical testing and clinical applications of aberration-free intraocular lenses
Visualitza/Obre
Estadístiques de LA Referencia / Recolecta
Inclou dades d'ús des de 2022
Cita com:
hdl:2099.1/9895
Tutor / directorHull, Cris C.
Tipus de documentProjecte Final de Màster Oficial
Data2010-09-09
Condicions d'accésAccés obert
Llevat que s'hi indiqui el contrari, els
continguts d'aquesta obra estan subjectes a la llicència de Creative Commons
:
Reconeixement-NoComercial-SenseObraDerivada 3.0 Espanya
Abstract
The spherical aberration of an Akreos ADAPT AO aberration-free intraocular
lens was studied using a point-diffraction interferometer and the ocular aberrations of
patients undergoing pre-operative assessment for implantation with this lens were
obtained with a Scheimpflug system (Pentacam) and Hartmann-Shack aberrometer
(Zywave). Measured values for the Z(4,0) coefficient were 0.028 μm +/- 0.019 and
0.021 μm +/- 0.013 for the Akreos lens and an equivalent spherically-surfaced
intraocular lens when measured in air with a collimated beam at 543.1nm. Anterior and
posterior corneal aberrations on 11 subjects demonstrated that the posterior cornea
contributed 1.2 times the spherical aberration of the anterior surface of the cornea. In
addition, lenticular spherical aberration values were found to be -1.543 μm +/- 0.29
when corneal aberration data was combined with whole eye aberrometry. In air
interferometry did not demonstrate lower values of spherical aberration for the aspheric
Akreos lens as expected. This may have been due to liquid film effects on the surfaces
negating the aspheric profile. Point-diffraction interferometry was found to be limited in
the amount of tilt that could be produced to generate tilt fringes for analysis. The
corneal posterior surface spherical aberration is not negligible and should be taken into
account when designing and selecting aspheric intraocular lenses. Customized
intraocular lenses would provide the optimal result for all patients.