Practical aspects of a pulse generator calibration

dc.contributor.authorHudlicka, Martin
dc.contributor.authorAzpúrua Auyanet, Marco Aurelio
dc.contributor.authorCelep, Murat
dc.date.accessioned2020-06-25T16:54:27Z
dc.date.available2020-06-25T16:54:27Z
dc.date.issued2020-04-13
dc.description© 2020 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes,creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works.
dc.description.abstractMeasuring receivers are used for measurement of radio disturbance in the frequency range typically 9 kHz to 18 GHz. Such receivers can be either electromagnetic interference (EMI) receivers or spectrum analyzers with the quasi-peak (QP) detector. Requirements for measuring receivers are discussed in European CISPR documents (Comité International Spécial des Perturbations Radioélectriques) and US standards ANSI 63.2 (QP parts derived from CISPR). Although novel techniques are being evaluated, the traditional way of checking compliance of the receiver with requirements of these standards is using a calibrated pulse generator. The calibration of pulse generators is discussed in the standard EN 55016-1-1, which is the harmonized version of the international standard IEC/ CISPR 16-1-1(currently Ed. 4). In the standard, however, only a very brief description of the methods is given and technical details are hidden. The measurement uncertainty of the pulse generator characterization is not discussed in the standards; however, it was discussed in several previous works. This paper aims to provide a more thorough description of particular calibration methods together with practical hints which may be useful for students, calibration engineers, and practitioners.
dc.description.peerreviewedPeer Reviewed
dc.description.sponsorshipThis work was supported by the project 15RPT01 RFMicrowave. This project has received funding from the EMPIR programme co-financed by the Participating States and from the European Union’s Horizon 2020 research and innovation programme
dc.description.versionPostprint (author's final draft)
dc.format.extent8 p.
dc.identifier.citationHudlicka, M.; Azpurua, M. A.; Celep, M. Practical aspects of a pulse generator calibration. "IEEE instrumentation and measurement magazine", 13 Abril 2020, vol. 23, núm. 2, p. 13-20.
dc.identifier.doi10.1109/MIM.2020.9062683
dc.identifier.issn1094-6969
dc.identifier.urihttps://hdl.handle.net/2117/191575
dc.language.isoeng
dc.relation.publisherversionhttps://ieeexplore.ieee.org/document/9062683
dc.rights.accessOpen Access
dc.subject.otherOscilloscopes
dc.subject.otherFrequency measurement
dc.subject.otherReceivers
dc.subject.otherPulse generation
dc.subject.otherGenerators
dc.subject.otherUncertainty
dc.subject.otherPulse measurements
dc.titlePractical aspects of a pulse generator calibration
dc.typeArticle
dspace.entity.typePublication
local.citation.authorHudlicka, M.; Azpurua, M. A.; Celep, M.
local.citation.endingPage20
local.citation.number2
local.citation.publicationNameIEEE instrumentation and measurement magazine
local.citation.startingPage13
local.citation.volume23
local.identifier.drac27743762