Design, control, and pilot study of a lightweight and modular robotic exoskeleton for walking assistance after spinal cord injury

dc.contributor.authorFont Llagunes, Josep Maria
dc.contributor.authorLugrís Armesto, Urbano
dc.contributor.authorClos Costa, Daniel
dc.contributor.authorAlonso Sánchez, Francisco Javier
dc.contributor.authorCuadrado Aranda, Javier
dc.contributor.groupUniversitat Politècnica de Catalunya. BIOMEC - Biomechanical Engineering Lab
dc.contributor.otherUniversitat Politècnica de Catalunya. Departament d'Enginyeria Mecànica
dc.contributor.otherUniversitat Politècnica de Catalunya. Doctorat en Enginyeria Mecànica, Fluids i Aeronàutica
dc.date.accessioned2020-02-17T11:45:31Z
dc.date.available2021-01-15T01:31:21Z
dc.date.issued2020-01-13
dc.description.abstractWalking rehabilitation using exoskeletons is of high importance to maximize independence and improve the general well-being of spinal cord injured subjects. We present the design and control of a lightweight and modular robotic exoskeleton to assist walking in spinal cord injured subjects who can control hip flexion, but lack control of knee and ankle muscles. The developed prototype consists of two robotic orthoses, which are powered by a motor-harmonic drive actuation system that controls knee flexion–extension. This actuation module is assembled on standard passive orthoses. Regarding the control, the stance-to-swing transition is detected using two inertial measurement units mounted on the tibial supports, and then the corresponding motor performs a predefined flexion–extension cycle that is personalized to the specific patient’s motor function. The system is portable by means of a backpack that contains an embedded computer board, the motor drivers, and the battery. A preliminary biomechanical evaluation of the gait-assistive device used by a female patient with incomplete spinal cord injury at T11 is presented. Results show an increase of gait speed (+24.11%), stride length (+7.41%), and cadence (+15.56%) when wearing the robotic orthoses compared with the case with passive orthoses. Conversely, a decrease of lateral displacement of the center of mass (-19.31%) and step width (-13.37% right step, -8.81% left step) are also observed, indicating gain of balance. The biomechanical assessment also reports an overall increase of gait symmetry when wearing the developed assistive device.
dc.description.peerreviewedPeer Reviewed
dc.description.versionPostprint (published version)
dc.format.extent8 p.
dc.identifier.citationFont-Llagunes, J.M. [et al.]. Design, control, and pilot study of a lightweight and modular robotic exoskeleton for walking assistance after spinal cord injury. "Journal of mechanisms and robotics", 13 Gener 2020, vol. 12, núm. 3, p. 1-8.
dc.identifier.doi10.1115/1.4045510
dc.identifier.issn1942-4302
dc.identifier.urihttps://hdl.handle.net/2117/177824
dc.language.isoeng
dc.publisherASME PRESS
dc.relation.projectidinfo:eu-repo/grantAgreement/MINECO//DPI2015-65959-C3-2-R/ES/ORTESIS HIBRIDA MOTOR-FES DE BAJO COSTE PARA LA MARCHA DE LESIONADOS MEDULARES Y METODOS DE SIMULACION PARA AYUDA AL DISEÑO Y LA ADAPTACION/
dc.relation.publisherversionhttps://asmedigitalcollection.asme.org/mechanismsrobotics/article-abstract/12/3/031008/1069010/Design-Control-and-Pilot-Study-of-a-Lightweight?redirectedFrom=fulltext
dc.rights.accessOpen Access
dc.rights.licensenameAttribution-NonCommercial-NoDerivs 3.0 Spain
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/3.0/es/
dc.subjectÀrees temàtiques de la UPC::Enginyeria biomèdica::Biomecànica
dc.subjectÀrees temàtiques de la UPC::Enginyeria biomèdica::Robòtica mèdica
dc.subject.lcshBiomechanics
dc.subject.lemacBiomecànica
dc.subject.otherMedical robotics
dc.subject.otherMulti-body dynamics and exoskeletons
dc.subject.otherWearable robots
dc.subject.otherRehabilitation engineering
dc.subject.otherBiomechanics
dc.subject.otherGait analysis
dc.subject.otherSpinal cord injury
dc.titleDesign, control, and pilot study of a lightweight and modular robotic exoskeleton for walking assistance after spinal cord injury
dc.typeArticle
dspace.entity.typePublication
local.citation.authorFont-Llagunes, J.M.; Lugrís, U.; Clos, D.; Alonso, F.J.; Cuadrado, J.
local.citation.endingPage8
local.citation.number3
local.citation.publicationNameJournal of mechanisms and robotics
local.citation.startingPage1
local.citation.volume12
local.identifier.drac26879292

Fitxers

Paquet original

Mostrant 1 - 2 de 2
Carregant...
Miniatura
Nom:
JMR-19-1118_final_authors.pdf
Mida:
3.56 MB
Format:
Adobe Portable Document Format
Carregant...
Miniatura
Nom:
JMR-19-1118_final_authors.pdf
Mida:
3.54 MB
Format:
Adobe Portable Document Format