A comprehensive model for power line interference in biopotential measurements
Títol de la revista
ISSN de la revista
Títol del volum
Col·laborador
Editor
Tribunal avaluador
Realitzat a/amb
Tipus de document
Data publicació
Editor
Condicions d'accés
item.page.rightslicense
Publicacions relacionades
Datasets relacionats
Projecte CCD
Abstract
Power line interference is a major problem in high-resolution biopotential measurements. Because interference coupling is mostly capacitive, shielding electrode leads and a high common-mode rejection ratio (CMRR) are quite effective in reducing power-line interference but do not completely eliminate it. We propose a model that includes both interference external to the measuring system and interference coming from its internal power supply. Moreover, the model considers interference directly coupled to the measuring electrodes, because, as opposed to connecting leads, electrodes are not usually shielded. Experimental results confirm that reducing interference coupled through electrodes yields a negligible interference. The proposed model can be applied to other differential measurement systems, particulary those involving electrodes or sensors placed far apart.