Tear film quality evaluation with intraocular scattering measurements
Visualitza/Obre
Estadístiques de LA Referencia / Recolecta
Inclou dades d'ús des de 2022
Cita com:
hdl:2117/86917
Tipus de documentProjecte Final de Màster Oficial
Data2010-08-28
Condicions d'accésAccés obert
Llevat que s'hi indiqui el contrari, els
continguts d'aquesta obra estan subjectes a la llicència de Creative Commons
:
Reconeixement-NoComercial-SenseObraDerivada 3.0 Espanya
Abstract
A new objective method to diagnose and classify dry eye patients has been
carried out in this study. Tear film quality was objectively evaluated by means of a
double-pass system in patients with healthy eyes, with some tear deficiency and dry eye
patients. Intraocular scattering present in those eyes was studied by analyzing a new
parameter called Objective Scatter Index (OSI) obtained with the Optical Quality
Analysis System (OQAS), a device based on the double-pass technique. Several
complementary subjective clinical exams commonly used to evaluate tear film quality
were also performed on patients (dry eye questionnaire, Non-Invasive Break-Up Time
(NIBUT), Phenol Red Thread Test and Visual Acuity) in order to correlate them with
the OSI parameter. Correlations between OSI and dry eye questionnaire and NIBUT
showed that there exists a trend that could help to classify more accurately patients
affected by some tear problem