Mostra el registre d'ítem simple

dc.contributor.authorAlbertini, M.
dc.contributor.authorHerrero Climent, Mariano
dc.contributor.authorLázaro, Pedro
dc.contributor.authorRíos, Jose Vicente
dc.contributor.authorGil Mur, Francisco Javier
dc.contributor.otherUniversitat Politècnica de Catalunya. Departament de Ciència dels Materials i Enginyeria Metal·lúrgica
dc.date.accessioned2012-10-05T11:34:15Z
dc.date.created2012-07-15
dc.date.issued2012-07-15
dc.identifier.citationAlbertini, M. [et al.]. Comparative study on AISI 440 and AISI 420B stainless steel for dental drill performance. "Materials letters", 15 Juliol 2012, vol. 79, p. 163-165.
dc.identifier.issn0167-577X
dc.identifier.urihttp://hdl.handle.net/2117/16661
dc.description.abstractOne of the most usual causes of the failure in dental implant osseointegration is the increase of the temperature above 47 °C during the drilling of the bone because it produces irreversible osteonecrosis. The result is weakened contact of implants with bone and possible loss of rigid fixation. Many studies about the design of the drill bit, contact forces, drill speed, drill point angle and feed-rate have been made. However, in all studies a martensitic stainless steel (AISI 440) has been used. The aim of this study was to determine the behavior of AISI420B stainless steel with higher hardness, higher heat conductivity and lower wear than the conventional stainless steel used in oral implantology. This steel showed adequate temperature increases, with or without irrigation, and a decrease of the weight lost of the drills with use.
dc.format.extent3 p.
dc.language.isoeng
dc.rightsAttribution-NonCommercial-NoDerivs 3.0 Spain
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/3.0/es/
dc.subjectÀrees temàtiques de la UPC::Enginyeria dels materials
dc.subject.lcshStainless stee
dc.subject.lcshBiomedical materials
dc.subject.otherBiomaterials
dc.subject.otherDrill
dc.subject.otherStainless steel
dc.subject.otherWear
dc.titleComparative study on AISI 440 and AISI 420B stainless steel for dental drill performance
dc.typeArticle
dc.subject.lemacAcer inoxidable
dc.subject.lemacBiomaterials
dc.contributor.groupUniversitat Politècnica de Catalunya. BBT - Biomaterials, Biomecànica i Enginyeria de Teixits
dc.identifier.doi10.1016/j.matlet.2012.04.006
dc.description.peerreviewedPeer Reviewed
dc.rights.accessRestricted access - publisher's policy
local.identifier.drac10614733
dc.description.versionPostprint (published version)
dc.date.lift10000-01-01
local.citation.authorAlbertini, M.; Herrero-Climent, M.; Lázaro, P.; Ríos, J.V.; Gil, F.J.