Mostra el registre d'ítem simple

dc.contributor.authorLázaro, Pedro
dc.contributor.authorHerrero Climent, Mariano
dc.contributor.authorGil Mur, Francisco Javier
dc.contributor.otherUniversitat Politècnica de Catalunya. Departament de Ciència dels Materials i Enginyeria Metal·lúrgica
dc.date.accessioned2012-02-15T11:24:46Z
dc.date.available2012-02-15T11:24:46Z
dc.date.created2010
dc.date.issued2010
dc.identifier.citationLázaro, P.; Herrero-Climent, M.; Gil, F.J. Evaluation of titanium dental implants after failure of osseointegration by means of X-ray photoelectron spectroscopy, electron microscopy and histological studies. "Journal of Biomedical Science and Engineering", 2010, vol. 3, núm. 11, p. 1073-1077.
dc.identifier.issn1423-0127
dc.identifier.urihttp://hdl.handle.net/2117/15147
dc.description.abstractIn this work, we analysed 56 clinically failed and retrieved implants by means of scanning electron microscopy (SEM), X-ray photoelectron spectroscopy (XPS) and histological studies. The surface contamination was compared to that of unused control implants and with that of the same implants after cleaning in a basic medium. The surfaces of the unused implants presented considerable contamination. In particular, high levels of carbon were detected. The nature of this C was elucidated by XPS analysis of the lubricant used in the machining process. The same contamination was observed in the retrieved implants. Histological studies were carried out by means of light microscopy. Fibrosis and granulomatous lesions were detected in the tissues. XPS analysis indicated the presence of traces of other elements (Na, Ca, Zn, S, F, etc.) that were not related to impurities in cpTi. We examined a cleaning process in a basic medium that eliminates the organic components of the implant surfaces. The cleaned implants were implanted in the patients and the results were excellent. None of the implants failed in following 7 months.
dc.format.extent5 p.
dc.language.isoeng
dc.publisherSpringer
dc.rightsAttribution-NonCommercial-NoDerivs 3.0 Spain
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/3.0/es/
dc.subjectÀrees temàtiques de la UPC::Enginyeria dels materials
dc.subject.lcshDental implants
dc.subject.lcshTitanium
dc.titleEvaluation of titanium dental implants after failure of osseointegration by means of X-ray photoelectron spectroscopy, electron microscopy and histological studies
dc.typeArticle
dc.subject.lemacImplants dentals
dc.subject.lemacTitani
dc.contributor.groupUniversitat Politècnica de Catalunya. BBT - Biomaterials, Biomecànica i Enginyeria de Teixits
dc.identifier.doi10.4236/jbise.2010.311139
dc.rights.accessOpen Access
local.identifier.drac9595409
dc.description.versionPostprint (published version)
local.citation.authorLázaro, P.; Herrero-Climent, M.; Gil, F.J.
local.citation.publicationNameJournal of Biomedical Science and Engineering
local.citation.volume3
local.citation.number11
local.citation.startingPage1073
local.citation.endingPage1077