The influence of the field setup on the dosimetry of abutted fields in single-isocenter half-beam techniques
Visualitza/Obre
The influence of the field.pdf (224,3Kb) (Accés restringit)
Sol·licita una còpia a l'autor
Què és aquest botó?
Aquest botó permet demanar una còpia d'un document restringit a l'autor. Es mostra quan:
- Disposem del correu electrònic de l'autor
- El document té una mida inferior a 20 Mb
- Es tracta d'un document d'accés restringit per decisió de l'autor o d'un document d'accés restringit per política de l'editorial
Cita com:
hdl:2117/12313
Tipus de documentArticle
Data publicació2011-03
Condicions d'accésAccés restringit per política de l'editorial
Llevat que s'hi indiqui el contrari, els
continguts d'aquesta obra estan subjectes a la llicència de Creative Commons
:
Reconeixement-NoComercial-SenseObraDerivada 3.0 Espanya
Abstract
Purpose: To study the influence of the field setup on the dosimetry at the junction in singleisocenter
half-beam techniques.
Methods: The dosimetry at the junction for a two-field setup with the gantry at zero was first
evaluated with radiochromic films. A three-field setup, with an anterior field and two opposed
lateral fields, was also analyzed for two different relative positions of the fields involved. In all
cases, the dose increase at the central axis, called the junction dose, was measured.
Results: Junction doses varied greatly with the setup. For the three-field setup, the junction dose
differed from that obtained with the two-field setup, and it greatly depended on the relative position
of the fields. When the anterior field was closer to the gantry than the lateral fields, a field gap
occurred and the junction dose was negative. When the anterior field was farther from the gantry
than the lateral fields, a field overlap was obtained and the junction dose was positive. The difference
in the junction dose between the three-field setups was around 18% for the three accelerators
evaluated.
Conclusions: Having a uniform dose distribution for two fields at gantry 0° does not guarantee a
uniform distribution at other gantry angles. Junction doses are largely affected by the relative
position of the radiation fields, which may have an impact in clinical practice. Therefore, any
method aiming to assess or to optimize the dose homogeneity at the junction should take this effect
into account.
CitacióHernández, V.; Sempau, J. The influence of the field setup on the dosimetry of abutted fields in single-isocenter half-beam techniques. "Medical physics", Març 2011, vol. 38, núm. 3, p. 1468-1472.
ISSN0094-2405
Fitxers | Descripció | Mida | Format | Visualitza |
---|---|---|---|---|
The influence of the field.pdf | 224,3Kb | Accés restringit |