| Citació: | Morgenstern, C.R. [et al.]. Automatic differentiation of obstructive and central hypopneas with esophageal pressure measurement during sleep. A: IEEE Engineering in Medicine and Biology Society. Conference. "31st Annual International Conference of the IEEE Engineering in Medicine and Biology Society". Minneapolis, Minnesota: 2009, p. 7102-7105. |
| Títol: | Automatic differentiation of obstructive and central hypopneas with esophageal pressure measurement during sleep |
| Autor: | Morgenstern de Muller, Christian Rudolf ; Schwaibold, Matthias; Randerath, Winfried J.; Bolz, Armin; Jané Campos, Raimon  |
| Data: | 2009 |
| Tipus de document: | Conference report |
| Resum: | The differentiation between obstructive and
central respiratory events is one of the most recurrent tasks in the diagnosis of sleep disordered breathing. Esophageal
pressure measurement is the gold-standard method to assess respiratory effort and identify these events. But as its
invasiveness discourages its use in clinical routine, noninvasisve systems have been proposed for differentiation. However, their adoption has been slow due to their limited
clinical validation, as the creation of manual, gold-standard validation sets by human experts is a cumbersome procedure.
In this study, a new system is proposed for an objective automatic, gold-standard differentiation between obstructive
and central hypopneas with the esophageal pressure signal.
First, an overall of 356 hypopneas of 16 patients were manually scored by a human expert to create a gold-standard validation
set. Then, features were extracted from each hypopnea to train and test classifiers (Discriminant Analysis, Support Vector
Machines and adaboost classifiers) to differentiate between central and obstructive hypopneas with the gold-standard
esophageal pressure signal. The automatic differentiation system achieved promising results, with a sensitivity of 0.88, a
specificity of 0.93 and an accuracy of 0.90. Hence, this system seems promising for an automatic, gold-standard differentiation between obstructive and central hypopneas. |
| ISBN: | 978-1-4244-3296-7/09 |
| URI: | http://hdl.handle.net/2117/7535 |
| Versió de l'editor: | 10.1109/IEMBS.2009.5332900 |
| Versió de l'editor: | http://ieeexplore.ieee.org/search/freesrchabstract.jsp?tp=&arnumber=5332900&queryText%3Dmorgenstern+c%26openedRefinements%3D*%26searchField%3DSearch+All |
| Apareix a les col·leccions: | Altres. Enviament des de DRAC Departament d'Enginyeria de Sistemes, Automàtica i Informàtica Industrial. Ponències/Comunicacions de congressos SISBIO - Senyals i Sistemes Biomèdics. Ponències/Comunicacions de congressos
|
| Comparteix: |
|
Queda prohibida la reproducció, transformació, distribució i comunicació pública d'aquesta obra. Es permet, en tot cas, la reproducció per a ús privat sempre i quan la còpia que se'n faci no sigui objecte d'utilització col·lectiva ni lucrativa (art. 31.2 del Reial Decret Legislatiu 1/1996, de 12 d'abril, pel qual s'aprova el Text Refós de la Llei de Propietat Intel·lectual, http://bibliotecnica.upc.es/sepi/legislacio.asp).
Per a qualsevol ús que es vulgui fer diferent al permès, dirigiu-vos a: sepi@upc.edu
|