| Citació: | Rosell, F.; Nescolarde, L.; Elvira, J. Total body water changes using segmental bioimpedance in healthy population with similar anthropometry. A: World Congress on Medical Physics and Biomedical Engineering. "IFMBE 2009 World Congress on Medical Physics and Biomedical Engineering". Munich: Springer Verlag, p. 60-63. |
| Títol: | Total body water changes using segmental bioimpedance in healthy population with similar anthropometry |
| Autor: | Rosell Ferrer, Francisco Javier ; Nescolarde Selva, Lexa Digna ; Elvira, J. |
| Editorial: | Springer Verlag |
| Data: | 2009 |
| Tipus de document: | Conference lecture |
| Resum: | Electrical bioimpedance was measured in 7 body segments, and also with the standard right-side method, using 11 electrodes at 6 different frequencies in 8 healthy male subjects with similar anthropometry. Our objective was to determine the capability of segmental bioimpedance measurements to estimate small changes of water on each segment (TWsegi)and total body water (TBW) in comparison with the standard right-side method. Water was also estimated with 40K and
DXA. Volunteers were measured before and after a 3.5% water load of their individual TBW. The expected TBW mean increment after water load was 1.45 l. The estimator with
lower Standard Error SE was the weight of the subject (0.15 l). For impedance methods, the SE of the segmental method was 0.94 l vs. 1.41 l for right-side. Segmental volume changes obtained by DXA and Impedance compared with expected values showed maximum differences of almost 2 l for DXA and 0.5 l
for Z in the abdomen. In conclusion, in a healthy sample with similar anthropometry, such as astronauts and athletes, the
use of a segmental impedance method improves the accuracy of the right-side method to estimate TBW. Changes in water segments estimated by impedance where more close to expected values than using DXA. |
| ISBN: | 978-3-642-03897-6 |
| URI: | http://hdl.handle.net/2117/6798 |
| Versió de l'editor: | http://www.wc2009.org/WORLD-CONGRESS-2009/IFMBE/Pages/default.aspx |
| Apareix a les col·leccions: | Altres. Enviament des de DRAC IEB - Instrumentació Electrònica i Biomèdica. Ponències/Comunicacions de congressos Departament d'Enginyeria Electrònica. Ponències/Comunicacions de congressos
|
| Comparteix: |
|
Queda prohibida la reproducció, transformació, distribució i comunicació pública d'aquesta obra. Es permet, en tot cas, la reproducció per a ús privat sempre i quan la còpia que se'n faci no sigui objecte d'utilització col·lectiva ni lucrativa (art. 31.2 del Reial Decret Legislatiu 1/1996, de 12 d'abril, pel qual s'aprova el Text Refós de la Llei de Propietat Intel·lectual, http://bibliotecnica.upc.es/sepi/legislacio.asp).
Per a qualsevol ús que es vulgui fer diferent al permès, dirigiu-vos a: sepi@upc.edu
|