Measurements of eye lens doses in interventional cardiology using OSL and electronic dosemeters
Visualitza/Obre
Cita com:
hdl:2117/26661
Tipus de documentArticle
Data publicació2014-12-01
Condicions d'accésAccés obert
Llevat que s'hi indiqui el contrari, els
continguts d'aquesta obra estan subjectes a la llicència de Creative Commons
:
Reconeixement-NoComercial-SenseObraDerivada 3.0 Espanya
Abstract
The purpose of this paper is to test the appropriateness of OSL and electronic dosemeters to estimate eye lens doses at interventional cardiology environment. Using TLD as reference detectors, personal dose equivalent was measured in phantoms and during clinical procedures. For phantom measurements, OSL dose values resulted in an average difference of 215% vs. TLD. Tests carried out with other electronic dosemeters revealed differences up to +/- 20% versus TLD. With dosemeters positioned outside the goggles and when TLD doses were > 20 mu Sv, the average difference OSL vs. TLD was 29%. Eye lens doses of almost 700 mu Sv per procedure were measured in two cases out of a sample of 33 measurements in individual clinical procedures, thus showing the risk of high exposure to the lenses of the eye when protection rules are not followed. The differences found between OSL and TLD are acceptable for the purpose and range of doses measured in the survey.
CitacióSanchez, R. [et al.]. Measurements of eye lens doses in interventional cardiology using OSL and electronic dosemeters. "Radiation protection dosimetry", 01 Desembre 2014, vol. 162, núm. 4, p. 569-576.