| Citació: | Fernandez Chimeno, M; Quilez, M; Silva, F. Understanding electrosurgical unit perturbations in order to address hospital operating room electromagnetic compatibility. IEEE Transactions on Biomedical Engineering, 2006 vol.53, núm. 6 p.1206-1209 |
| Títol: | Understanding electrosurgical unit perturbations in order to address hospital operating room electromagnetic compatibility |
| Autor: | Fernández Chimeno, Mireya ; Quílez Figuerola, Marcos ; Silva Martínez, Fernando  |
| Editorial: | IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC |
| Data: | 30-jun-2006 |
| Tipus de document: | Article |
| Resum: | We analyze the radiated emissions from a low power electrosurgical unit (ESU). Measurements at a 3-meter distance are performed in order to find a suitable measurement setup intended to reproduce realistically the real working ESU behavior in a test lab. The broad-band and narrow-band characteristics of the perturbation are studied in order to address the other equipment immunity. |
| ISSN: | 0018-9294 |
| URI: | http://hdl.handle.net/2117/1346 |
| Apareix a les col·leccions: | Departament d'Enginyeria Electrònica. Articles de revista IEB - Instrumentació Electrònica i Biomèdica. Articles de revista
|
| Comparteix: |
|
Queda prohibida la reproducció, transformació, distribució i comunicació pública d'aquesta obra. Es permet, en tot cas, la reproducció per a ús privat sempre i quan la còpia que se'n faci no sigui objecte d'utilització col·lectiva ni lucrativa (art. 31.2 del Reial Decret Legislatiu 1/1996, de 12 d'abril, pel qual s'aprova el Text Refós de la Llei de Propietat Intel·lectual, http://bibliotecnica.upc.es/sepi/legislacio.asp).
Per a qualsevol ús que es vulgui fer diferent al permès, dirigiu-vos a: sepi@upc.edu
|