E-prints UPC >
Altres >
Enviament des de DRAC >
Empreu aquest identificador per citar o enllaçar aquest ítem:
http://hdl.handle.net/2117/12637
|
Ítem no disponible en accés obert per política de l'editorial
| Arxiu |
Descripció |
Mida | Format |
| perfomance.pdf | | 176.14 kB | Adobe PDF |  |
|
| Citació: | Ginjaume, M. Performance and approval procedures for active personal dosemeters. "Radiation protection dosimetry", Març 2011, vol. 144, núm. 1-4, p. 144-149. |
| Títol: | Performance and approval procedures for active personal dosemeters |
| Autor: | Ginjaume Egido, Mercè  |
| Data: | mar-2011 |
| Tipus de document: | Article |
| Resum: | Active personal dosemeters (APDs) are well accepted as useful and reliable instruments for individual dosimetry measurements.
The increasing concern about studying the behaviour of APDs in pulsed fields is illustrated through revision of the
results of the most representative studies on the performance of APDs in the last 5 y. The deficiencies of APDs in pulsed
fields are discussed together with proposals to overcome them. Although there are no legal constraints or technical limitations
for recognising APDs for legal dosimetry in facilities with continuous radiation fields, APDs continue to be mainly used as
operational dosemeters. The approval procedures applicable to APDs, especially the approach undertaken by Germany, are
presented. Finally, some trends in the developments and use of APDs are summarised. |
| ISSN: | 0144-8420 |
| URI: | http://hdl.handle.net/2117/12637 |
| Versió de l'editor: | 10.1093/rpd/ncq457 |
| Apareix a les col·leccions: | Altres. Enviament des de DRAC GREENER - Grup de recerca d´estudis energètics i de les radiacions. Articles de revista
|
| Comparteix: |
|
Aquest ítem (excepte textos i imatges no creats per l'autor) està subjecte a una llicència de Creative Commons Llicència Creative Commons
|