| Angel_sdarticle-39_Fabrication...revMicroelectornic-Elsev.pdf | text complet | 779.21 kB | Adobe PDF |  |
|
| Citació: | Valera, E.; Muñiz, D.; Rodriguez, A. Fabrication of flexible interdigitated µ-electrodes (FIDµEs) for the development of a conductimetric immunosensor for atrazine detection based on antibodies labelled with gold nanoparticles. "Microelectronic engineering", 2010, vol. 87, núm. 2, p. 167-173. |
| Títol: | Fabrication of flexible interdigitated µ-electrodes (FIDµEs) for the development of a conductimetric immunosensor for atrazine detection based on antibodies labelled with gold nanoparticles |
| Autor: | Valera Cano, Enrique Andrés ; Muñiz, David; Rodríguez Martínez, Ángel  |
| Data: | 2010 |
| Tipus de document: | Article |
| Resum: | The present paper describes a simple and low-cost method for the fabrication of mechanically flexible interdigitated μ-electrodes (FIDμEs) and its application as immunosensor.
FIDμEs consist of two coplanar non-passivated interdigitated metallic μ-electrodes supported on a flexible transparent substrate, polyethylene naphthalate (PEN). Bioreagents deposition on PEN substrates becomes possible by depositing SiO2 on the electrodes surface (fingers and inter-digits space).
These FIDμEs were successfully applied for the development of a selective conductimetric immunosensor for the quantification of atrazine residues. The immunosensor has been demonstrated for detection of small amounts of atrazine, thanks to the use of immunoreagents specifically developed to detect this pesticide.
The detection method applied is based on the use of antibodies labelled with gold nanoparticles. The presence of these particles amplifies the conductive signal; hence the immunosensor response was quantified using simple and inexpensive DC measurements.
Immunochemical detection of the concentrations of atrazine is achieved by a competitive reaction which occurs before the inclusion of the labelled antibodies.
The immunosensor shows limits of detection in the order of 2–3 μg L−1, far below the maximum residue level (50 μg kg−1) established by EU for residues of atrazine as herbicide in the wine grapes and other foodstuff products. |
| ISSN: | 0167-9317 |
| URI: | http://hdl.handle.net/2117/12582 |
| Versió de l'editor: | 10.1016/j.mee.2009.07.001 |
| Versió de l'editor: | http://www.sciencedirect.com/science?_ob=ArticleURL&_udi=B6V0W-4WR19X5-1&_user=1517299&_coverDate=02%2F28%2F2010&_rdoc=1&_fmt=high&_orig=gateway&_origin=gateway&_sort=d&_docanchor=&view=c&_searchStrId=1673202402&_rerunOrigin=google&_acct=C000053450&_version=1&_urlVersion=0&_userid=1517299&md5=222cc783d1b090ed59133d8a259c80e3&searchtype=a |
| Apareix a les col·leccions: | Altres. Enviament des de DRAC Departament d'Enginyeria Electrònica. Articles de revista MNT - Grup de Recerca en Micro i Nanotecnologies. Articles de revista
|
| Comparteix: |
|
Queda prohibida la reproducció, transformació, distribució i comunicació pública d'aquesta obra. Es permet, en tot cas, la reproducció per a ús privat sempre i quan la còpia que se'n faci no sigui objecte d'utilització col·lectiva ni lucrativa (art. 31.2 del Reial Decret Legislatiu 1/1996, de 12 d'abril, pel qual s'aprova el Text Refós de la Llei de Propietat Intel·lectual, http://bibliotecnica.upc.es/sepi/legislacio.asp).
Per a qualsevol ús que es vulgui fer diferent al permès, dirigiu-vos a: sepi@upc.edu
|